Preface
This is the third edition of CSA Z314.15, Storage, transportation, and distribution of single use and reusable medical devices. It supersedes the previous edition published in 2010. It is one of the CSA Z314 series of Standards dealing with decontamination, disinfection, sterilization and handling of sterile medical devices.
This Standard specifies requirements for the storage, handling, transportation, and distribution of single use and reusable medical devices in facilities where open-inventory (split-case) storage and handling services are provided.
This Standard is intended to assist health care settings and other storage facilities in establishing safe and effective storage and handling practices to ensure that medical devices are protected from microbial contamination and other damage. This Standard specifies a systems approach, recognizing the trend in health care to use shared, centralized storage solutions that can be located on and/or off site and introduces classification categories (tiers) for storage facilities.
This Standard has been revised to work in tandem with applicable requirements of CSA Z314.0, Medical device reprocessing – General requirements, and other standards in the CSA Z314 series.
CSA group would like to acknowledge the contributions of Andrée Pelletier, Johanne Dionne, Diane Pinsonnault, and Anne Marie Rancourt towards the development of this Standard.
This Standard was prepared by the Subcommittee on Warehousing of Medical Supplies, under the jurisdiction of the Technical Committee on Sterilization and the Strategic Steering Committee on Health Care Technology Systems.
Scope
1.1
This Standard specifies requirements and recommended practices for the storage, handling, transportation, and distribution of single use and reusable medical devices in facilities where open-inventory (split-case) storage and handling services are provided.
1.2
This Standard covers
a) physical and functional requirements for storerooms, and other storage facilities that provide open-inventory (split-case) storage of single use and reusable single use or reusable medical devices;
Note: See CSA Z314.0 for additional requirements for sterile storage areas for reprocessed products.
b) environmental conditions for storage and transportation;
c) staff qualifications, orientation, and education; occupational health and safety; and other personnel considerations;
d) storage and handling of single use and reusable medical devices;
e) management of case cart systems;
f) onsite distribution; and
g) emergency procedures.
1.3
The storage facilities covered by this Standard include the following:
a) sterile storage areas associated with operating room and medical device reprocessing areas including third party/off site reprocessors (tier 1) (i.e., single use opened case products and reprocessed medical devices);
b) storage areas outside of the sterile core within the health care setting that store both clean and sterile inventory (tier 2) (i.e., single use opened case products and reprocessed medical device); and
c) storage areas that distribute and/or transport open single use inventory clearly separated from storage areas that distribute and/or transport closed cases (tier 3).
Note: Tier 3 can include onsite and offsite warehouse storage locations.
1.4
This Standard does not apply to
a) the reprocessing of reusable medical devices that have become contaminated through accident or mishandling during storage, transportation, or on-site distribution;
Note: See CSA Z314.8.
b) the storage of medical devices at the point of manufacturing;
c) transportation between the point of manufacturing and the warehouse or storage area; and
d) warehousing that does not include low unit of measure (opened case product) distribution (e.g., vendor warehouse with bulk storage).
1.5
In this Standard, “shall” is used to express a requirement, i.e., a provision that the user is obliged to satisfy in order to comply with the standard; “should” is used to express a recommendation or that which is advised but not required; and “may” is used to express an option or that which is permissible within the limits of the Standard.
Notes accompanying clauses do not include requirements or alternative requirements; the purpose of a note accompanying a clause is to separate from the text explanatory or informative material.
Notes to tables and figures are considered part of the table or figure and may be written as requirements.
Annexes are designated normative (mandatory) or informative (non-mandatory) to define their application.
1.6
The values given in SI units are the units of record for the purposes of this Standard. The values given in parentheses are for information and comparison only.